Frequency Load range Dimensions |
0,7 to 300 Hz (active isolation), passive above 300 Hz |
Isolation: |
active 0.7 Hz to 300 Hz, |
Transmissibility: |
see attached curve Above 10 Hz trans- missibility < 0,01 (-40dB) |
Correction Forces: |
+/- 8 N vertical |
Static Compliance: |
approx. 12 µm/N vertical |
Maximum Load: |
150 kg |
Size: |
400 x 450 x 78.5 mm |
Weight: |
17,5 kg |
Electrical |
|
Safety class: |
1 |
Power Consumption: |
maximum 10W |
Input Voltage: |
115 - 230 V AC +/- 10% 50 - 60 Hz |
Overload: |
The display indicates when the load is too heavy or severely unbalanced |
Monitor signal: |
A multiplexed signal for display on oscilloscope shows vibration levels with and without isolation - for diagnostic purposes only |
Front panel Display menu: |
Instrumental status |
Table top |
|
Size: |
400 x 450 mm |
Material: |
Damped Aluminium sandwich construction |
Download attachments: